ACTIVA CLIMATIC (in complianche with EN-12341)


Activa Climatic cabinet has been developed to allow filters conditioning at 20 ± 1°C and at 50 ± 5 %RH, before and after use.

The internal dimensions of the cabin are such to allow the permanent housing of an analytical balance and a congruous numbers of filters.

The cabin is realized in accordante with the following requirements:
UNI EN 12341 (April 2001)
Method UNICHIM n° 285 (ed. 2003)
D. M. n° 60,  2 April 2002


- Chassis: cohibented plate
- Frontal panel: in polycarbonate, with opening flap, with two portholes (Ø 150 mm)
- Lighting: 11 watt fluorescent lamp
- Display: backlighted LCD
- Operating signals: acoustic signal and  LED luminous
- Printer: included, for recording data (serial number of the cab - date - time - UR - T). The data of hour, RH and T  are recorded every 15  minutes (the operator can modify this range, depending on his needs, from 1 to 25 min).
- USB output: included. Hour, RH and  T, if the cabin is connected to a PC by USB output, can be downloaded for the development of graphs illustrating the evolution of T and RH during the entire period of operation of the cab.
- Power supply: 230 V
- External dimensions: 1020 x 890 x 935 mm ( l x p x h)  
- Internal dimensions: 
                                      width: 920 mm
                                      depth: 600 mm

                                      frontal height: 590 mm
                                      back height: 500 mm 
- Weight: 60 Kg

Limitations of installation: the cabinet must be placed in a room where the microclimatic conditions are between  18 to 26°C  and 40 to 65% RH (different conditions could compromise the cabinet operation).