ACTIVA CLIMATIC (in complianche with EN-12341)
|
Activa Climatic cabinet has been developed to allow filters conditioning at 20 ± 1°C and at 50 ± 5 %RH, before and after use.
The internal dimensions of the cabin are such to allow the permanent housing of an analytical balance and a congruous numbers of filters.
The cabin is realized in accordante with the following requirements: UNI EN 12341 (April 2001) Method UNICHIM n° 285 (ed. 2003) D. M. n° 60, 2 April 2002 |
|
- Chassis: cohibented plate - Frontal panel: in polycarbonate, with opening flap, with two portholes (Ø 150 mm) - Lighting: 11 watt fluorescent lamp - Display: backlighted LCD - Operating signals: acoustic signal and LED luminous - Printer: included, for recording data (serial number of the cab - date - time - UR - T). The data of hour, RH and T are recorded every 15 minutes (the operator can modify this range, depending on his needs, from 1 to 25 min). - USB output: included. Hour, RH and T, if the cabin is connected to a PC by USB output, can be downloaded for the development of graphs illustrating the evolution of T and RH during the entire period of operation of the cab. - Power supply: 230 V - External dimensions: 1020 x 890 x 935 mm ( l x p x h) - Internal dimensions: width: 920 mm depth: 600 mm frontal height: 590 mm back height: 500 mm - Weight: 60 Kg
Limitations of installation: the cabinet must be placed in a room where the microclimatic conditions are between 18 to 26°C and 40 to 65% RH (different conditions could compromise the cabinet operation). |
|