mini FLOW_FLV


The cabinets of the series mini FLOW have been developed for the protection of operator, product and environment, with the criterion of the typical vertical laminar flow cabinets (Class II non Biohazard), but smaller.

The air of downflow, decontaminated from the absolute filter HEPA, goes with laminar flow on the work area, crosses the worktop and mixes with the outside air that enters the cabin through the front opening, in the front area of the worktop. The contaminated air is aspirated from the fan in the upper part of the cabinet and conveyed in a plenum: here about 70% is sent back into the working area after absolute filtration and the remaining 30% is just expelled outside (mod. FLV), or expelled after absolute filtration (mod. FLV/H) or absolute filtration and filtration on molecular carbon filter (mod. FLV/HC). The quantity of expelled air is reintegrated with an equal amount of ambient air aspirated through the front opening, creating a curtain to protect the operator.

The cabinets are versatile and, due to their small size, easily locatable when required (even using their special support with wheels).


- External chassie: steel sheet painted in epoxy powder stove

- Double back panel for intake to even the horizontal air flow

- Worktop: stainless steel inox AISI 304 perforated, made by two panels removable for cleaning

- Painted sheet metal work tray for liquid recovery under the worktop

- Frontal window in temperate glass (thickness 5 mm) openable with the possibility of locking in the fully open position

- Height front opening: min 200 mm – max 350 mm

- Thermally protected motor fan 

- Exhaust collar (diameter: 150 mm) for possible conveyance outside of the exhaust air through PVC pipe

- Fluorescent lamp in the area outside the work chamber (power 13 W)

- Illumination work area: > 700 lux

- UVC germicidal lamp: 15 W power

Control pannel in the front of machine with “soft touch” membrane keyboard

- High quality electronic board controlled by a microprocessor

- Digital hour meter of the operating of the machine and of the germicidal UV lamp (if installed)

- Interlock fluorescent lamp/UV lamp

- Countdown time  meter  for UV lamp programmable by the customer with automatic switch off

- Early warning visual on display of exceeding the maximum period for the replacement of filters (resettable)

- Capacity:  350 m3/h

- Average air speed: 0,4 m/sec

- Class work: 100 (second ISO5)

- Possibility of manual adjustment of the airflow by the operator directly on the keyboard to:

   * eventually modify the standard factory setting for particular business needs

   * gradually increase the voltage of motor fan to avoid the progressive clogging of the pre-filters and HEPA filters, if installed

- Noise level:  < 58 dB(A)

- Power installed:  250 W   

- Electric power:  230V - 50 Hz

- External dimensions:  800 x 580 x 940 mm (L x W x H)

- Internal useful dimensionse:  795 x 500 x 575 mm (L x W x H)

- Net weight:

   ° mini Flow FLV:      70 kg
   ° mini Flow FLV_H:   73 kg
   ° mini Flow FLV_HC: 78 kg
- Compliance: built in compliance with:

   * ISO 5 (ex Federal Standard 209D)

   * Electrotechnical Italian Committee (CEI 66.5)
   * CEI 62.25 Norms for Class 1 fixed installation equipment
   * UNI-CIG